HP 8642A Operation And Calibration Manual page 277

Synthesized signal generator
Hide thumbs Also See for 8642A:
Table of Contents

Advertisement

Model 8642A/B
Performance Tests
PERFORMANCE TESTS
4-11. INTERNAL MODULATION OSCILLATOR TESTS
SPECIFICATION:
DESCRIPTION:
EQUIPMENT:
Electrical Characteristics
Performance Limits
Conditions
FREQUENCY
Range
20 Hz to 100 kHz
Resolution
1 % of setting
Accuracy
2% of setting
OUTPUT LEVEL
Range
o
to 3 volts peak
600 ohm load
Resolution
4 mV
Accuracy
±(4% off setting +15 rnV)
within 1 second
Distortion
<0.02%
0.02-15.8 kHz
(>0.5 volt peak)
<0.15%
>15.8 kHz
Output Impedance
600 ohms ± 10%
All Internal Modulation Oscillator specifications are checked by measuring
the specified parameters on an audio analyz.er.
HP 8642A/B SYNTHESIZED
SIGNAL GENERATOR
o
0
0
I
10
D
IS 8 88 BB88B
0
0
[5
D
0 0 0 0 0
l5!5 EifsEHs8
0
~
a
AUDIO
ANALYZER
I
·
INPUT
L -_ _ _ _ _ ... -
_~~E!.1_
-- - - ••
f
STEP 5:
:
LITJ-.i
600 OHM
FEEDTHROUOH
TERMINATION
Figure 4-7. Internal Modulation Oscillator Test Setup
Audio Analyz.er ........................ HP 8903A
Feedthrough Termination, 600 ohm ... HP 1109 SA
(a 600 ohm 1% resistor to ground
can be substituted. Use HP part number
0757-1100or 0698-4015.)