Model 8642A/B
Performance Tests
PERFORMANCE TESTS
4-13. SPURIOUS TEST
SPECIFICA TION:
DESCRIPTION:
EQUIPMENT:
PROCEDURE:
Electrical Characteristics
Performance Limits
ConditiDns
Non harmonic spurious
100 dBc
band HET-9
94 dBc
band 10
The RF output of the HP 8642 is viewed on a spectrum analyzer. The
8642's frequency is set to values where spurious performance problems are
most likely to be observed.
10 MHz
o
0
0
I
10
OUTPUT
000
~D~
§§ §§§§§
0
~D
RF OUTPUT
HP 8642A/B SYNTHESIZED
SIGNAL GENERATOR
EXTERNAL
REFERENCE
SPECTRUM
RF INPUT
ANALYZER
Figure
4-9.
Harmonics and Spurious Test Setup
Spectrum Analyzer ................... HP 8566A
1.
Perform a center frequency calibration on the spectrum analyzer.
2.
Connect the instruments as shown in Figure 4-9.
3.
Set up the spectrum analyzer as follows:
INSTRUMENT PRESET
SPAN
0 kHz
RESOLUTION BANDWIDTH 10Hz
4.
Set up the HP 8642 as follows:
INSTRUMENT PRESET