Getting the Best Results

Success in obtaining accurate and consistent test results will depend on the
care with which test procedures are carried out. Always follow the test
instructions carefully and observe the stated standing periods and temperature
conditions where applicable.
Wipe test tubes free from condensation before placing in the photometer. Test
tubes should always be kept in a clean condition. Wash and dry carefully after
use. Dirty tubes may be soaked in weak detergent solution if necessary.
Tubes which become stained or scratched in use should be replaced.

Taking Test Readings

1 Press
key. The instrument displays the 'Choose a Test' menu box, with
the last test program used highlighted as the active line.
The cursor will flash on the [OK] symbol of the 'options menu' at the bottom
of the screen.
Press [OK] to accept this program.
2 To choose a different test program, either use the
through the menu options, or use the numeric keys to enter the Phot
number of the desired test.
Press [OK] to accept the selected program.
3 The following display will appear :-
Place a BLANK TUBE in the test chamber, then press [OK].
NOTE: The instrument is designed to hold the blank setting as long as the instrument
is switched on. This stage will be omitted when further tests are being carried out.
However, when changing to a test which requires a coloured or reagent blank, or
uses a tube of a different diameter, the 'Insert Blank' prompt will be displayed.
4 The instrument will be set automatically, and after a few seconds the
following display will appear :-
Place SAMPLE TUBE in the test chamber, then press [OK].
Phot 007
Chlorine-Free / 5
Insert Blank
OK
Choose a Test
Phot 007
Chlorine-Free / 5
Insert Sample
OK
Blank - Choose a Test - Time
and
keys to scroll
8