Remove and disassemble the printer assembly
Separate the top housing assembly from the bottom housing assembly as described.
1. Remove the ground cable at the J603 connection from the printer board (PCA).
2. Remove the four M4x7 screws from the printer assembly using a # 2 Phillips bit.
Reassembly notes
Use a #2 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure
the printer assembly M4x7 screws.
Route the ground cable around the outside of the screw boss during reassembly.
Disassembly, repair, and reassembly 57