56 Disassembly, repair, and reassembly
Remove the power supply assembly
Separate the top housing assembly from the bottom housing assembly as described.
1. Remove the P1 and J1 connectors from the power supply assembly.
2. Remove the four 4-20 self-tapping screws from the power assembly using a #1
Phillips bit.
3. Remove the power supply.
Reassembly note
Use a #1 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure the
power supply screws.
CP 150 12-lead resting electrocardiograph