Welch Allyn CP 150 Service Manual page 25

12-lead resting electrocardiograph
Hide thumbs Also See for CP 150:
Table of Contents

Advertisement

Service manual
USB host test
Note
1. Insert a USB flash drive into each of the four ports on the back of the device and
follow the instructions on the screen to test each of the ports.
2. Touch the USB host test button to test each port from left to right.
3. Touch the
USB client test
1. Connect a USB type mini-B connector to the device client USB port and a type A
connector to a free USB port on the PC. Verify that the PC is able to recognize
CP150 device.
2. Touch the Yes button to confirm that your PC recognized the CP 150 device, or
touch the No button if your PC did not recognized the CP 150 device.
3. Touch the
Welch Allyn has not validated specific USB flash drives. Use only 64Gb or
less USB flash drives.
(Next) button to perform the next functional test.
(Next) button to perform the next functional test.
Service menu 21