Example:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLD <Time>
This command defines the demodulation time at the marker position.
Prerequisites for this command
Turn on a marker (CALCulate<n>:MARKer<m>[:STATe]).
Turn on marker demodulation
DEModulation[:STATe]).
Select a frequency span > 0 ([SENSe:]FREQuency:SPAN).
For measurement in the time domain (span = 0), the command has no effect,
because demodulation is continuous.
Suffix:
<n>
<m>
Parameters:
<Time>
Example:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect <Demodulation>
This command selects the demodulation type for marker demodulation.
Prerequisites for this command
Turn on a marker (CALCulate<n>:MARKer<m>[:STATe]).
Turn on marker demodulation
DEModulation[:STATe]).
User Manual 1178.4130.02 ─ 02
北京海洋兴业科技股份有限公司(证券代码:839145)
Remote Commands of the Spectrum Application
//Turn on single sweep mode and marker 1.
INIT:CONT OFF
CALC:MARK ON
//Turn on the frequency counter for marker 1.
CALC:MARK:COUN ON
//Perform a measurement and query the results of the frequency
counter.
INIT;*WAI
CALC:MARK:COUN:FREQ?
(CALCulate<n>:MARKer<m>:FUNCtion:
.
irrelevant
irrelevant
<numeric value>
Range:
100 ms to 500 s
*RST:
500 ms
Default unit: s
//Configure marker demodulation
FREQ:SPAN 100MHZ
CALC:MARK ON
CALC:MARK:X 100MHZ
CALC:MARK:FUNC:DEM ON
CALC:MARK:FUNC:DEM:SEL FM
CALC:MARK:FUNC:DEM:HOLD 3s