<n>
<m>
Parameters:
<DisplayMode>
Example:
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:UNIT <Unit>
This command selects the unit of the for channel power measurements.
Prerequisites for this command
Select the channel power measurement
FUNCtion:POWer:SELect).
Suffix:
<n>
<m>
Parameters:
<Unit>
Example:
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ <State>
This command turns the display of the channel power in a 1 Hz bandwidth on and off.
Prerequisites for this command
Select the channel power measurement
FUNCtion:POWer:SELect).
Suffix:
<n>
<m>
Parameters:
<State>
Example:
User Manual 1178.4130.02 ─ 02
北京海洋兴业科技股份有限公司(证券代码:839145)
Remote Commands of the Spectrum Application
.
irrelevant
irrelevant
CLR
Display the currently measured value ("Clear / Write").
MAX
Display the highest measured value ("Max Hold").
*RST:
CLR
//Select clear/write trace mode
CALC:MARK:FUNC:CPOW:MODE CLR
(CALCulate<n>:MARKer<m>:
.
irrelevant
Marker
DBM | DBMV | DBUV | VOLT | WATT | V | W
*RST:
DBM
//Select the unit for the channel power
CALC:MARK:FUNC:CPOW:UNIT DBM
(CALCulate<n>:MARKer<m>:
.
irrelevant
irrelevant
ON | OFF | 1 | 0
*RST:
OFF | 0
//Display the measured channel power in a 1 Hz bandwidth
CALC:MARK:FUNC:POW:RES:PHZ ON
Remote Control
184