Maintenance - Miele PG 8527 Operating And Installation Instructions

Laboratory glassware washer
Table of Contents

Advertisement

Maintenance

Labwashers should undergo routine
maintenance by the Miele Technical
Service every 1000 operating hours
or at least once every 6 months.
Maintenance covers the following:
– electrical safety
– Door mechanism and door seal
– Screws and connections in the wash
cabinet
– Water inlet and drain
– Internal and external dispensing
systems
– Spray arms
– Triple filter system
– Drain pump and non-return valve
– Steam condensor
– All mobile units, baskets, and inserts
If applicable:
– Drying unit
– Connected printer
38
Also during maintenance, the following
function checks are conducted:
– a test program is run
– wash temperatures are measured
– ensure a water tight door seal
– all safety-relevant measuring
systems are checked (fault display).

Hide quick links: