Interrupting A Program - Miele PG 8527 Operating And Installation Instructions

Laboratory glassware washer
Table of Contents

Advertisement

Interrupting a program

A program can only be interrupted
or cancelled in operating levels B
and D.
Only interrupt a program if the door
must be opened for urgent reasons,
e.g., if items are moving around too
much, or the cleaning performance
needs to be checked.
If a program with process statement
is interrupted then continued, make
sure to check the display upon
program completion. If the display
Process parameter not
reads
achieved
, the door was opened
after process parameter monitoring
began, and so the process
parameter was not achieved. The
program may need to be repeated.
In operating levels B or D
^ Press the +C button.
The program is interrupted.
The following prompts appear in the
display:
Program Cancel (OK)
or Continue (Clear)?
^ Open the door a
Note PG 8528:
Which door can be opened depends
on which setting has been programmed
under "Machine Function/Door
Mechanism" (see the Programming
Manual).
The door on the unclean side can
always be opened, no matter which
parameter has been set.
The door on the clean side can only be
opened if one of the following
parameters are set:
- Sluice,
- Door clean + Sluice,
- Door unclean + Sluice.
,
Caution! Items in the machine
may be hot. There is the danger of
burning or scalding.
^ Arrange items securely.
^ Close the door a.
^ Press the +C button.
The program continues.
Operation