Maintenance
Baskets and inserts
To ensure that the mobile units,
baskets, and inserts function properly,
they must be checked routinely. A
checklist is supplied with the
labwasher.
Check the following points:
– Are the rollers in proper condition,
and are they securely attached to the
basket / insert?
– Are the counter-nuts on the guide
rails firmly secured?
– Is the basket connection correctly
set and firmly screwed on?
– For baskets in the modular system,
are the caps in the module
connection working properly?
– Are all spray jets, spray sleevs, and
hose adapters securely attached to
the basket / insert?
– Are all spray jets, spray sleeves, and
hose adapters unclogged so that
wash water can flow through?
– Are all caps, covers, and fasteners
securely attached to the spray
sleeves?
44
If there are spray arms:
– Do the spray arms rotate freely?
– Are the spray jets clogged? See the
section on "Cleaning and Care /
Cleaning the spray arms".
– Are the screws in the magnet rails for
Automatic Mobile Unit Recognition
tightly secured?
– Are the magnets in place on the
spray arm ends?
– Are the magnets on the spray arms
free of attached metal parts?