Main Switch - On/Off
The main switch disconnects the
labwasher from power supply
^ Turn the main switch to I-ON.
Once the start procedure is completed,
the labwasher is ready to for use.
The following message may appear:
Fault 2218: "New power frequency
60 Hz Flow rate automatically
adjusted" or "New power frequency
50 Hz Flow rate automatically
adjusted".
During the start procedure, the
Profitronic system detected a
different power frequency, and
adjusted the flow rate of the
dispensing pump accordingly.
Before operation, you will need to
check if the change in power
frequency was caused by a fault in
the power supply.
If so, incorrect amounts of
processing chemicals would be
dispensed during the program.
The fault message can only be
confirmed in operating level D and
in the service mode.
Switching On
^ Press and hold the s button .
The activating time for the s button
can be set to a maximum of 10
seconds (see "System functions
/Actuating time ON/OFF Button" in
the Programming Manual).
Depending on the set operating level,
the following will appear in the display:
Operating
Displayed
levels
A and B
A list of approved
programs.
C
Instructions for
Automatic Mobile Unit
recognition.
D
Select from:
– Program Overview
– Programming
– Settings J.
If the washer is not in use, the
display automatically switches off
after approx. 10 min.
To turn the display back on, touch
any button.