Cleaning The Spray Arms - Miele PG 8527 Operating And Installation Instructions

Laboratory glassware washer
Table of Contents

Advertisement

Maintenance

Cleaning the spray arms

The spray arms can become clogged.
Check daily and clean if needed.
^ Use a sharp, pointed object to push
any particles into the spray arm jets,
then rinse throughly under running
water.
To do so, remove the spray arms as
follows:
^ Remove the basket.
Labwasher spray arms
^ Loosen the nut with a wrench and
pull the spray arm either upward or
downward.
42
Spray arms on basket / inserts
^ Use a wrench to loosen the nut on
the spray arm guide, then pull the
spray arm downward.
If you notice wear and tear on the
bearings or the gasket, please
contact Miele.
^ After cleaning, put the spray arms
back in place and tighten the nut.
Make sure the nut is in the proper
position.
After replacing the spray arms make
sure that they rotate freely.
,
Make sure that no metallic parts
or pieces become attached to the
magnets at the ends of the spray
arms.
These objects can cause the spray
arm rotations to be miscalculated.