MC20
variable. This is always displayed also in the main menu or context menu
for constant visual control.
7.3.2
Defined Interval Measurement
The measurement over a defined period of time is the standard interval
measurement. It is used, for example, for series measurements with a
constant time interval.
Setting
Trigger/end measurement
22
Status: Profile Interval measurement
1. Press
in profile
2. Call Interval menu item, confirm with
3. Enter duration with
Setting range: 0-9999 s (max. 166min:39s).
Input 0 s deactivates the interval function and
then only momentary values are shown in the
Interval measurement profile too. This is
indicated by the missing prefixed Mean in the
names of the measuring variables!
•
The interval is now defined. The measurement is
triggered in the main menu with START Interval.
•
End running intervals prematurely with STOP
Interval in the main menu.
•
Display jumps back to the profile for the value
view after pressing START Interval or STOP
Interval.
Main Menu
.
, confirm with
Schiltknecht Messtechnik AG
.