MEAN VALUE MEASUREMENT
Schiltknecht Messtechnik AG
Instantaneous values may not be informative enough under strongly
fluctuating conditions.
ManoAir100 therefore provides two ways for mean value readings
(damping).
A)
standard (preset) averaging time of 6 secs.
B)
individual time setting from 2s...2h..
Mean value is displayed after lapse of such time.
A)
Standard 6-secs averaging time
With pitot tube brought to the measuring point.
Press <Rep> key.
Display shows countdown of seconds from 6 to 0, followed by mean
value reading.
If pressing of <Rep> key is followed immediately by pressing
key, the continuous values are displayed with MEAN sign blinking
on and off.
Countdown is performed internally. Mean value reading is dis-
played afterwards.
ManoAir100
23