Enter Programming In Advanced Mode - Honeywell AreaRAE Plus User Manual

Gas
Hide thumbs Also See for AreaRAE Plus:
Table of Contents

Advertisement

AreaRAE Plus & AreaRAE Pro User's Guide
Once you enter Programming Mode, the Calibration menu is highlighted. You cannot access other menu
items, and you can only perform the following types of calibration:
Fresh Air
Multi Sensor Span
Single Sensor Zero
Single Sensor Span
The following require password access in Basic Mode or can be directly accessed in Advanced
Mode.
Multi Sensor Bump
Single Sensor Bump
Cal. Reference
Change Cal. Gas
Multi Cal Select
Change Span Value
To enter a menu and view or edit parameters in its submenus, press [Y/+].

12.2 Enter Programming In Advanced Mode

1. To enter Programming Mode, press and hold [MODE] and [N/-] until you see the Calibration screen.
No password is necessary in Advanced Mode.
2. Press [N/-] to step through the programming screens.
Note: Holding down [N/-] advances quickly through the menu options.
When a menu item is selected, its name is shown at the top of the screen, and its icon is enlarged.
To enter a menu and view or edit parameters in its submenus, press [Y/+].
37