AreaRAE Plus & AreaRAE Pro User's Guide
Single Sensor Bump
This menu allows a bump test to be performed on an individual sensor of your choice.
Note: If a bump test icon (bottle with bottom portion not filled in) is shown next to any of the sensors, it
means that the sensor is due for a bump test.
To perform a bump test on an individual sensor, follow these steps:
1. At the Calibration Menu, select "Single Sensor Bump."
2. Scroll down the list using [N/-], and then press [Y/+] to select a sensor to bump test.
3. Install the "T" tube and connect it to a source of calibration gas.
4. Verify that the displayed calibration value meets the concentration specified on the gas cylinder.
5. Start the flow of calibration gas.
Note: Dotted line indicates automatic progression.
6. Press [Y/+] to start calibrating or wait for the bump test to start automatically.
7. A countdown screen appears. You can abort the bump test at any time during the countdown by
pressing [N/-].
8. If the bump test is not aborted, the display shows the sensor names and tells you whether the bump
test passed or failed, followed by the sensor readings.
Important!
After each sensor is bump tested, and you press
"OK," the next sensor in the menu list is highlighted.