AreaRAE Plus & AreaRAE Pro User's Guide
3. Press [Y/+] to select.
4. Press [Y/+] to "Save." You can also press [N/-] to undo.
Zero At Start
If your AreaRAE Plus/Pro has been configured to perform a zero (fresh air) calibration upon startup,
called Zero At Start, then the startup routine is interrupted so that you can perform a fresh air calibration
for all sensors prior to using the instrument.
1. Press [Y/+] to see on or off state, and which is selected.
2. Press [N/-] to scroll to "On" or "Off."
3. Press [Y/+] to select.
4. Press [Y/+] to "Save." You can also press [N/-] to undo.
If you do not want to perform a zero calibration, press [MODE] to bypass it. If you start a zero calibration
and want to abort it, press [N/-], and the calibration stops and the main display is shown.
Fast Startup
Fast Startup reduces the amount of time between when the instrument is turned on and is ready for use. It skips
showing you many settings and is best suited to environments where the AreaRAE Plus/Pro is turned on and
off very often during a given day. If Fast Startup is not selected, then when the instrument starts, it shows you
details of each sensor, including calibration information, high and low alarm settings, etc.
1. Press [Y/+] to see on or off state, and which is selected.
2. Press [N/-] to scroll to "On" or "Off."
3. Press [Y/+] to select.
4. Press [Y/+] to "Save." You can also press [N/-] to undo.
Language
English is the default language, but other languages can also be selected for the instrument. Note: The
language can also be changed through ProRAE Studio II.
1. Press [Y/+] to see the current, selected language.
2. Press [N/-] to scroll to the language you want.
3. Press [Y/+] to select.
4. Press [Y/+] to "Save." You can also press [N/-] to undo.
Site ID
Choose and enter an 8-digit Site ID to uniquely identify the particular site where the instrument is to be
used. The first four digits can be an alphabet letter or number, while the last four digits can only be
numbers. This Site ID is included in the datalog report.
Note: Advance through the alphabet and numbers (0 through 9) by one with each press of the [Y/+] key.
To scroll quickly, hold down the [Y/+] key for as long as you want it to scroll rapidly.
User ID
Enter an 8-digit alphanumeric User ID to uniquely identify a user. This User ID is included in the datalog
report. The first four characters of a customized User ID act as an identifier for montoring the instrument.
Note: Advance through the alphabet and numbers (0 through 9) by one with each press of the [Y/+] key.
To scroll quickly, hold down the [Y/+] key for as long as you want it to scroll rapidly.
53