AreaRAE Plus & AreaRAE Pro User's Guide
8. If the calibration is not aborted, the display shows the sensor names and tells you whether the
calibration passed or failed, followed by the sensor readings.
Note: The gamma radiation sensor comes pre-calibrated from the factory and does not require routine
calibration. However, you can check it by placing a check-source on the rear of the AreaRAE Plus/Pro
equipped with a gamma sensor to check the readings. Place the check-source near the upper middle of the
rear of the instrument and hold it until you see the Gamma reading change in the display.
Multi Sensor Bump
Depending on the configuration of your AreaRAE Plus/Pro and the span gas you have, you can perform a
bump test simultaneously on multiple sensors.
The selected sensors are shown on the screen. With calibration gas connected to the instrument, start a
multiple bump test by applying gas to the instrument and pressing [Y/+]. If you do not want to perform a
multiple bump test, press [MODE].
Note: You can abort a multiple bump test by pressing [N/-] once testing has started.
When the Multi Bump test is done, a screen is shown, with the sensor names and either "Pass" or "Fail"
shown next to them.