INSTRUCTION MANUAL
3561
3561-01
BATTERY HiTESTER

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3561 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hioki 3561

  • Page 1 INSTRUCTION MANUAL 3561 3561-01 BATTERY HiTESTER...
  • Page 4 Contents Chapter 4 Applied Measurement _______________________ 33 Comparator Function..........34 4.1.1 Comparator Setting Example 1 (Upper and Lower Threshold Judgment) ...... 35 4.1.2 Comparator Setting Example 2 (Reference Value and Tolerance Judgment) ....39 4.1.3 Comparator Judgment Beeper Setting ......43 4.1.4 Comparator Execution Mode Setting ......
  • Page 5 RS-232C/GP-IB Interfaces ____________________ 87 Overview and Features ..........87 Specifications ..............88 7.2.1 RS-232C Specifications ..........88 7.2.2 GP-IB Specifications (Model 3561-01 only) ....88 Selecting the Connections and Protocol ....89 7.3.1 Attaching the Connector ..........89 7.3.2 Selecting the Interface ..........91 Communication Methods ..........92 7.4.1...
  • Page 6 Contents Chapter 8 Specifications _____________________________ 157 Basic Specifications..........157 Accuracy..............161 General Specifications..........162 Chapter 9 Maintenance and Service ___________________ 163 Troubleshooting............163 Cleaning ..............164 Error Display............. 164 Appendix _________________________________ 165 Appendix 1 Precautions for Making Custom Test Leads .. 165 Appendix 2 AC Four-terminal Method .......
  • Page 38 3.7 Displaying Measurement Results 3.7.1 Measurement Fault Detection If a measurement does not execute properly, a measurement fault “- - - - -” is indicated on the display. In addition, a measurement fault signal (ERR) is output at the EXT I/O connector.
  • Page 74 4.12 Reset Function...
  • Page 78: Output Signals

    5.2 Signal Descriptions 0ADJ Zero adjustment executes once when the signal transitions from 0ADJ High to Low. PRINT The current measurement value prints when the signal PRINT transitions from High to Low. MANU When the MANU comparator mode is selected, comparator judgment is enabled while the signal is Low.
  • Page 82 5.3 Timing Chart Description Time 1.5 ms ERR Output response time Measurement trigger 0.5 ms min. pulse width Delay Time per setting See Section 4.2.2 Trigger Delay Settings (Page 52). Ω Ω Measurement time V mode mode or V mode EX.FAST 6.8 ms EX.FAST...
  • Page 92 6.4 Printing Example Printouts _____________________________________________ Measurement values Measurement values Measurement values (ΩV mode) (Ω mode) (V mode) 298.60mOhm, 1.3924 V 209.98mOhm 3132 4.2019 V 0.2984 Ohm, 1.3924 V 0.2103 Ohm 3133 15.2084 V 3.35mOhm, 0.0000 V 4 - 0.0054 Ohm, 0.0000 V 299.10mOhm,- 1.3923 V 0.2984 Ohm,- 1.3923 V...
  • Page 100 7.4 Communication Methods Message This instrument recognizes the following message terminators: Terminators • LF • CR • CR+LF • CR+LF • EOI • LF with EOI From the instrument's interface settings, the following can be selected as the terminator for response messages. •...
  • Page 106 7.4 Communication Methods 7.4.4 Event Registers Standard Event Status Register (SESR) The Standard Event Status Register is an 8-bit register. If any bit in the Standard Event Status Register is set to 1 (after masking by the Standard Event Status Enable Register), bit 5 (ESB) of the Status Byte Register is set to 1.
  • Page 112 7.5 Message List Message ([ ] = optional) Data Contents Description ( ) = response data page Comparator OFF/ HL/ IN/ BOTH1 Sets the comparator judgment beeper :CALCulate:LIMit:BEEPer / BOTH2 setting OFF/ HL/ IN/ BOTH1 Queries the comparator judgment :CALCulate:LIMit:BEEPer? / BOTH2 beeper setting HL/ REF...
  • Page 114 7.5 Message List Message ([ ] = optional) Data Contents Description ( ) = response data page Line Frequency 50/ 60 Selects the AC line frequency :SYSTem:LFRequency 50/ 60 Queries the AC line frequency :SYSTem:LFRequency? selection Key-Lock 1/ 0/ ON/ OFF Sets the key-lock :SYSTem:KLOCk ON/ OFF...
  • Page 118 ∗ Note command is supported because it is defined in IEEE 488.2- 1987, but because all Model 3561(3561-01) device-specific commands are sequential types, this command has no actual affect. Status and Event Control Commands Clear the Status Byte and Related Queues (Except the Output Queue)
  • Page 120 7.6 Message Reference Set and Query the Service Request Enable Register (SRER) ∗ Syntax Command <0 to 255> ∗ Query SRE? Response <0 to 255 (NR1)> Description Command The SRER mask is set to the numerical value 0 to 255. Although NRf numerical values are accepted, values to the right of the decimal are rounded to the nearest integer.
  • Page 122 7.6 Message Reference Select and Query the Measurement Mode Setting Syntax Command :FUNCtion <RV/ RESistance/ VOLTage> Query :FUNCtion? Response <RV/ RESISTANCE/ VOLTAGE> RV......ΩV mode (Resistance and voltage measurement) RESISTANCE ..Ω mode (Resistance measurement) VOLTAGE ..V mode (Voltage measurement) Example Command :FUNC RV...
  • Page 124 7.6 Message Reference Set and Query the Averaging Function Setting Syntax Command :CALCulate:AVERage:STATe <1, 0, ON or OFF> Query :CALCulate:AVERage:STATe? Response <ON or OFF> Example Command :CALC:AVER:STAT OFF Query :CALC:AVER:STAT? Response Set and Query the No. of samples to average Syntax Command :CALCulate:AVERage...
  • Page 126 7.6 Message Reference Set and Query the Comparator Upper Threshold Setting (Resistance Measurement) Syntax Command :CALCulate:LIMit:RESistance:UPPer <Upper threshold> Query :CALCulate:LIMit:RESistance:UPPer? Response <Upper threshold> = 0 to 99999 <Upper threshold> (NR1) Command Example :CALC:LIM:RES:UPP 28593 Sets the upper threshold to 285.93 mΩ (with the 300 mΩ range selected) (If the 3 Ω...
  • Page 128 7.6 Message Reference Set and Query the Comparator Reference Value (Resistance Measurement) Syntax Command :CALCulate:LIMit:RESistance:REFerence <Reference value> Query :CALCulate:LIMit:RESistance:REFerence? Response <Reference value> = 0 to 99999 <Reference value> (NR1) Example Command :CALC:LIM:RES:REF 5076 Sets the reference value to mΩ (with the 300 mΩ range 50.76 selected) Ω...
  • Page 130 7.6 Message Reference Query Comparator Judgment Results (Resistance Measurement) Syntax Query :CALCulate:LIMit:RESistance:RESult? Response <HI/ IN/ LO/ OFF/ ERR> Example Query :CALC:LIM:RES:RES? Response (Voltage Measurement) Syntax Query :CALCulate:LIMit:VOLTage:RESult? Response <HI/ IN/ LO/ OFF/ ERR> Execute Statistical Functions Syntax Command :CALCulate:STATistics:STATe <1, 0, ON or OFF> Query :CALCulate:STATistics:STATe? Response...
  • Page 132 7.6 Message Reference Query the Maximum value (Resistance Measurement) Syntax Query :CALCulate:STATistics:RESistance:MAXimum? Response <Maximum value (NR3)>,<Data No. of Maximum value (NR1)> Example Query :CALC:STAT:RES:MAX? Response 297.28E-3,15 (Voltage Measurement) Syntax Query :CALCulate:STATistics:VOLTage:MAXimum? Response <Maximum value (NR3)>,<Data No. of Maximum value (NR1)> Example Query :CALC:STAT:VOLT:MAX?
  • Page 134 7.6 Message Reference Query the Process Capability Indices (Resistance Measurement) Syntax Query :CALCulate:STATistics:RESistance:CP? Response <Cp (NR2)>,<CpK (NR2)> Example Query :CALC:STAT:RES:CP? Response 0.04, 0.04 (Voltage Measurement) Syntax Query :CALCulate:STATistics:VOLTage:CP? Response <Cp (NR2)>,<CpK (NR2)> Example Query :CALC:STAT:VOLT:CP? Response 0.91, 0.00 Set and Query the Memory Function State Syntax Command :MEMory:STATe...
  • Page 136 7.6 Message Reference Execute Self-Calibration Syntax Command :SYSTem:CALibration Self-Calibration State and Setting Command :SYSTem:CALibration:AUTO <1, 0, ON or OFF> Query :SYSTem:CALibration:AUTO? Response <ON or OFF> ON... AUTO Self-Calibration selected (executes approximately every 30 minutes) OFF MANUAL Self-Calibration selected Example Command :SYST:CAL:AUTO ON Query :SYST:CAL:AUTO?
  • Page 138 7.6 Message Reference Set and Query EXT I/O Lock Syntax Command :SYSTem:ELOCk <1, 0, ON or OFF> Query :SYSTem:ELOCk? Response <ON or OFF> ON..EXT I/O control is disabled (preventing inadvertent operations from electrical noise). OFF..EXT I/O control is enabled. Example Command :SYST:ELOC ON...
  • Page 140 7.6 Message Reference System Reset Syntax Command :SYSTem:RESet Description Command All settings including saved panel settings are returned to factory defaults. Refer to “Reset Function” for details. Example Command :SYST:RES ∗ Note • If you want to preserve saved data, use the command instead.
  • Page 144 7.6 Message Reference Set and Query the Trigger Source Syntax Command :TRIGger:SOURce <IMMediate/ EXTernal> Query :TRIGger:SOURce? Response <IMMEDIATE/ EXTERNAL> IMMEDIATE .. Internal triggering EXTERNAL ... External trigger source. Triggering by key, TRIG ∗TRG TRIG terminal or command. Description Command Selects the trigger source. Query Queries the trigger source selection.
  • Page 146 7.6 Message Reference Execute a Measurement and Read the Measured Values Syntax Query :READ? Response <Measured resistance (NR3)>, <Measured voltage (NR3)> Ω V mode) Ω <Measured resistance (NR3)> mode) <Measured voltage (NR3)> (V mode) Description Query Switches from the Idle State to the Trigger Wait State, then reads the next measured value.
  • Page 148 7.6 Message Reference 7.6.4 Command Compatibility with the Model 3560 AC mΩ HiTESTER Model 3561 and 3561-01 BATTERY HiTESTERs accept all of the commands supported by the HIOKI 3560 AC mΩ HiTESTER. However the following differences result from the functional differences.
  • Page 150 7.6 Message Reference Message ([ ] = optional) Data Contents Differences ( ) = response data Model 3561 (3561-01) Model 3560 <Upper threshold/ Lower Setting range: 0 to 3.1000E+0 Setting range: 0 to 3.1000E+3 :CSET:RPARameter threshold> (only valid within 300 mΩ and 3 :CSET:RPARameter? Ω...
  • Page 154 7.8 Sample Programs (2) Measure Resistance by PC Key Measures and imports by key input on the PC, and saves measurements in a text file. Private Sub MeasureReadSubRS() Dim recvstr As String 'Receiving char string Dim i As Integer MSComm1.Settings = "9600,n,8,1" 'Comm port setting MSComm1.PortOpen = True 'Open a port...
  • Page 156 7.8 Sample Programs (4) External Trigger Measurement 2 Measure and import according to external triggering of the instrument ( key or EXT I/O TRIG TRIG terminal input), and save measurements in a text file. (The instrument imports the most recent measurement by trigger input timing with the continuous measurement state) Private Sub MeasureTrig2SubRS() Dim recvstr As String...
  • Page 158 Dim ud As Integer 'State (unused) Dim i As Integer pad = 0 'Board Address 0 gpibad = 1 '3561 (3561-01) Address 1 timeout = T10s 'Timeout about 10s Call ibfind("gpib0", 0) 'Initialize GP-IB Call ibdev(pad, gpibad, 0, timeout, 1, 0, ud) Call SendIFC(pad) Open App.Path &...
  • Page 160 'State (unused) im i As Integer pad = 0 'Board Address 0 gpibad = 1 '3561 (3561-01) Address 1 timeout = T100s 'Timeout 100s (because of external trigger wait state) Call ibfind("gpib0", 0) 'Initialize GP-IB Call ibdev(pad, gpibad, 0, timeout, 1, 0, ud) Call SendIFC(pad) Open App.Path &...
  • Page 162 'Timeout period Dim ud As Integer 'State (unused) pad = 0 'Board Address 0 gpibad = 1 '3561 (3561-01) Address 1 timeout = T10s 'Timeout about 10s Call ibfind("gpib0", 0) 'Initialize GP-IB Call ibdev(pad, gpibad, 0, timeout, 1, 0, ud) Call SendIFC(pad) 'Select ΩV mode...
  • Page 164: Response Time

    8.1 Basic Specifications Sampling Time Sampling rate EX.FAST/ FAST/ MEDIUM/ SLOW (four steps) Sampling time Sampling EX.FAST FAST MEDIUM SLOW ΩV (50 Hz) 83 ms 258 ms 7 ms 23 ms (60 Hz) 69 ms 252 ms Ω (50 Hz) 42 ms 157 ms 4 ms...
  • Page 166 Output to printer via RS-232C (multi-use) Printer : Model 9670 (option) Communications settings: Data length (8 bit), stop bit (1 bit), parity (none) Baud rate : 9600 bps GP-IB Applicable GP-IB Standards: IEEE488.2 (Model 3561-01 only) Address : 0 to 30 Delimiter : LF/ CR+LF...
  • Page 182 Appendix 9 Dimensional Diagram...
  • Page 184 Index ................ 72 Noise ........4, 6, 132, 163, 166 INDEX ..............72 ..............95 Initialize .............. 66 ..............95 Input buffer ............97 ..............95 Input impedance ..........157 Input terminal ............9 Inspection ............21 .............. 32, 157 Interface ............
  • Page 185 Index Sample program ..........147 Sampling .............63 Sampling rate ..........11, 28 Sampling time ...........158 Self calibration ..........63, 161 SENSE-H ............9, 165 SENSE-L ............9, 165 Separator ............94 SOURCE-H ...........9, 165 SOURCE-L ...........9, 165 Stand ..............12 Standard command ........104, 111 Standard deviation of population ......54 Statistical calculation ...........11...
  • Page 186 Index...
  • Page 189 HIOKI 3561, 3561-01 BATTERY HiTESTER Instruction Manual Publication date: September 2006 Revised edition 1 Edited and published by HIOKI E.E. CORPORATION Technical Support Section All inquiries to International Sales and Marketing Department 81 Koizumi, Ueda, Nagano, 386-1192, Japan TEL: +81-268-28-0562 / FAX: +81-268-28-0568 E-mail: os-com@hioki.co.jp...
  • Page 190 HEAD OFFICE 81 Koizumi, Ueda, Nagano 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp / URL http://www.hioki.co.jp/ HIOKI USA CORPORATION 6 Corporate Drive, Cranbury, NJ 08512, USA TEL +1-609-409-9109 / FAX +1-609-409-9108 3561A981-01 06-09H Printed on recycled paper...