9 Diagnostics and service
Failure
38
still valid. Plan in maintenance for the instrument because a failure is
expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
The following table shows the error codes and text messages in the
status message "Failure" and provides information on causes as well
as corrective measures.
Example for a failure message
Code
Cause
Text mes-
sage
F008
– Additional sensors not
switched on
Error multi
– EMC influences
sensor com-
– No other sensor available
munication
F013
– Error on the current input
– No valid measured value
Sensor sig-
– Connected instruments
nals a fault
without function
F016
– Values of the min. and max.
adjustment exchanged
Adjustment
data ex-
changed
F017
– The values of the min. and
max. adjustment are too
Adjustment
close together
span too small
F025
– Wrong value in the lineari-
zation table
Invalid lineari-
zation table
F030
– Process values are not
within the adjusted measur-
Process value
ing range
out of limits
F034
– Electronics defective
EPROM hard-
ware error
F035
– Error in the internal instru-
ment communication
EPROM data
error
F036
– Error during software
update
Faulty pro-
gram memory
Rectification
– Check wiring between the
sensors
– Connect the sensors cor-
rectly and make them ready
for operation
– Check current input
– Check connected instru-
ments (Slaves)
– Correct adjustment data
– Correct adjustment data
– Correct linearization table
– Repeat adjustment
– Exchanging the electronics
– Carry out a reset
– Exchanging the electronics
– Repeat software update
– Exchanging the electronics
POINTRAC 31 • 8/16 mA/HART - four-wire