Specifications; General Specifications - Fluke 1550 User Manual

Megohmmeter
Hide thumbs Also See for 1550:
Table of Contents

Advertisement

1550
Users Manual

Specifications

General Specifications

Display
Power
Charger Input (AC)
Charger Input (DC)
W
Guard Fuse
W
Charger Fuse
Dimensions (H x W x L)
Weight
Temperature (operating)
Temperature (storage)
Humidity
Altitude
Enclosure Sealing
Input Overload Protection
Electromagnetic Compatibility
Certifications
Safety Compliance
Typical Battery Charge
Capability
Note
At higher temperatures, the
battery requires charging more
frequently.
24
75 mm x 105 mm
12 V lead-acid rechargeable battery.
Yuasa NP2.8-12
85 V to 250 V ac 50/60 Hz 50 VA
10 to 15 V dc, 30 VA
HBC 100 mA, 250 V, Type F (Fast) to IEC127
5 A, 1.5-in x 0.25-in quick blow (F) fuse
170 mm x 242 mm x 330 mm
(6.7 in. x 9.5 in. x 13.0 in.)
4 Kg (8.8 lbs.)
−20
o
o
C to 40
C (−4
−20
o
o
C to 65
C (−4
o
80 % to 31
C decreasing linearly to
o
50 % at 40
C
2000 m
IP40
600 V to earth ground
EN 61326
UL, CE, CSA, TÜV
EN 61010, EN 61557 Parts 1 and 2
IEC 61010-1, CAT III V 600,
Pollution Degree 2
Test Voltages
500 V
1 kV
2.5 kV
5 kV
o
o
F to 104
F)
o
o
F to 149
F)
Number of Tests
3600
3200
2500
1500

Hide quick links:

Advertisement

Table of Contents
loading