ABB Sensyflow iG Operating Manual page 54

Intelligent gas-mass flowmeter
Table of Contents

Advertisement

Type
Operating pressure
Operating temperature of
medium
(Transducer)
Ambient temperature:
Evaluation electronics
Without display
With display
Storage temperature
Degree of protection
Recommended
installation requirements
Pressure loss
(logarithmic diagram)
Electrical power values
Power dissipation
Connections
Cable (remote version)
Output signals
PROFIBUS DPV1 version
Analog/HART version
Analog output
Digital output
Digital input
54
Sensyflow iG, Version 1.00
Sensyflow iG
Standard range: -25...+150 °C
Extended range: -25...+300 °C
minimum inlet run 15 × pipe diameter D, outlet run 5 × pipe diameter D (see page 11)
< 1.0 kPa (10 mbar), typical value 0.1 kPa (1 mbar)
10
5
1
0,5
0,1
10
20 VA
Round signal line with braided copper shield LIYCY 10 x 0.5 mm
max. 25 m between transducer and electronics unit
2 x passive optocoupler (approx. 100 mA) selectable as frequency, pulse, or switching output
2 x 24 V lin typ. 10 mA (low < 2 mA, high > 10 mA) frequency and switching input
Type overview, technical data
Sensyflow iG-Ex
4 × 10
6
Pa (40 bar)
-25...+65 °C
-25...+50 °C
Other ambient temperatures on request
-25...+85 °C
IP 67 (IP 65 for remote version transducer)
According to DIN EN ISO 5167-1
DN 25
DN 50
50
100
500
1000
Universal power supply unit:
110...230 V AC/DC ± 10%(f = 48...62 Hz)
Low-voltage power supply unit:
24 V AC/DC ± 20%(f = 48...62 Hz)
M20 x 1.5 or ½" NPT
EN 50170, acc. to PA profile 3.0
0/4...20 mA, load < 600 Ω
In preparation
DN 100
DN 80
DN 150
5000 10000
15 VA
2
.
42/14-39 EN