Precise Instrument DAT 400 Technical Manual page 30

Digital transmitter analog/profibus/devicenet versions
Table of Contents

Advertisement

FUN
FUN
ParaN
SET
SET
PRG
f1LtEr
0
Select Value
PRG
SET
FUN
Not1on
0
Select Value
PRG
SET
FUN
auto 0
0
Enter Value
PRG
SET
FUN
0 trac
0
Select Value
PRG
SET
FUN
0 band
0
Enter Value
PRG
SET
FUN
0 ALL
0
Select Value
PRG
SET
FUN
Page 28
0 trAc
In-oUt
This function allows you to perform a momentary zero calibration
compensating for the temperature drift of the weight.
At swithching off, it automatically returns to the previous zero cali-
PRG
bration.
The maximum weight resettable by this parameter is 2% of the range
PRG
of the system.
To disable this feature, use the value 0.
PRG
Value
PRG
0
1
PRG
2
3
PRG
4
Default: 0
0 band
This parameter defines the number of divisions resettable by the
pressure of the front button of zero or by Input 1.
Values: from 0 to 200
Default: 100
0 ALL ZERO CAL
Enables the function that allows to perform zero calibration by pres-
sing and holding the 0 key or input 1 for 5 sec.
Value:
Off
On
Default: Off
TRACKING THE ZERO
Change
Control OFF
0.5 div/sec
1 div/sec
2 div/sec
3 div/sec
ZERO BAND
Function disabled
Function Enabled