The parameters in this menu allow the adjusment of the acquisition and updating timing of the display
and the manual or automatic zeroing performed by transmitter
WEIGHT FILTER
F1LtEr
This parameter adjusts the refresh speed of the display and the serial
and analog output.
Low values of the filter speed up the display refresh.
High values of the filter slow down the display refresh.
Value
Update
0
50 Hz
1
50 Hz
2
25 Hz
3
25 Hz
4
25 Hz
5
10 Hz
6
10 Hz
7
10 Hz
8
5 Hz
9
5 Hz
Default: 5
WEIGHT STABILITY
Not1on
This parameter defines the divisions number needed to deem the
weight stable.
A large number of divisions allows the transmitter to detect quickly
the weight stability, which is needed when executing tare and print
commands.
Value
Change
0
Always stable weight
1
Stability determined quickly
2
Stability determined with medium parameters
3
Stability determined accurately
4
Stability determined with the highest accuracy
Default: 2
AUTOZERO UPON SWITCHING ON
Auto 0
This parameter defines the value of the maximum resettable weight
when the instrument is switched on.
This operation corresponds to a zero calibration of the system and
is executed only if the weight is stable and below the set value.
Value from 0 to the value of the CAPAC parameter.
Default: 0
WEIGHING PARAMETERS
Response
25 Hz
16 Hz
8 Hz
5 Hz
2.5 Hz
1.5 Hz
1 Hz
0.7 Hz
0.4 Hz
0.2 Hz
FUN
CaL1br
ParaN
SET
f1LtEr
0
SET
FUN
Not1on
0
SET
FUN
auto 0
0
SET
FUN
0 trac
0
SET
FUN
0 band
0
SET
FUN
0 ALL
0
SET
FUN
FUN
In-oUt
SET
PRG
Select Value
PRG
PRG
Select Value
PRG
PRG
Enter Value
PRG
PRG
Select Value
PRG
PRG
Enter Value
PRG
PRG
Select Value
PRG
PRG