Installation

Read the Bath Use section before placing the bath in service.
Incorrect handling can damage the bath and void the warranty.
This bath is not designed to be portable. Therefore, moving the bath once it has
been installed should be kept to a minimum.
Never move a bath that is full of fluid. This action could be
extremely dangerous and could result in personal injury to the
person moving the bath.
If the bath is going to be placed in an area where it will need to be moved
frequently, Fluke Calibration sells a special cart designed to accommodate the
bath. Thus making the bath much more portable. However, even with a cart the
bath should not be moved full of "hot" fluid. The fluid can splash causing injury or
if the bath and cart tip, the "hot" fluid could cause damage to the surrounding
area and personal injury to personnel.
Care must be used when moving the bath with the fluid
solidified in the bath. The control probe can easily be damaged.
If the bath must be moved, all the salt to solidify and cool. To safely move the
bath, two people are required. One person should lift the bath by the tower, while
the second person slides their hands underneath the bath tank and lifts the bath.

Bath Environment

DO NOT place under a cabinet or other structure. Allow for
overhead clearance.
The 6050H bath is a precision instrument which should be located in an
appropriate environment. The location should be free of drafts, extreme
temperatures and temperature changes, dirt, etc. The surface where the bath is
placed must be level.
Because the bath is designed for operation at high temperatures, keep all
flammable and meltable materials away from the bath. Although the bath is well
insulated, top surfaces do become hot. Beware of the danger of accidental fluid
spills. We recommend placing the bath on a heat-proof surface such as concrete
with plenty of clear space around the
If the bath is operated at high temperatures, a fume hood should be used to
remove any vapors given off by hot bath fluid.
WCaution
WWarning
WCaution
WCaution
bath.
Calibration Bath
Installation
13