18 Advanced Options for DPOAE Testing

NOTE: The Advanced Options for DPOAE Testing section is reserved for those units purchased
NOTE:
NOTE:
NOTE:
as a diagnostic device. Instruments purchased as a screener can skip to Section 20.
The Advanced Options menu permit modification of the test parameters and pass criterion
for the customizable DP protocols. Changes to the protocol should be made only by qualified
personnel, usually the administrator. If you are not familiar with the use of these variables, do
not attempt to change the protocols. Changes to any of these characteristics may yield test
results that differ from those obtained in other test modes.
®
The ERO•SCAN
instrument comes with pre-programmed protocol settings. See Appendix D
for the manufacturer settings of these protocols. Test protocol changes are saved in the non-
volatile memory so the settings will be retained even when the battery is discharged
temporarily.
Display
Display 31
31
Display
Display
31
31
18.1

18.1 Instructions for Customizing a Test Protocol

18.1
18.1
Instructions for Customizing a Test Protocol
Instructions for Customizing a Test Protocol
Instructions for Customizing a Test Protocol
To enter the DPOAE Menu:
1) Press
2) Using the     CHANGE
customize (the "DP 4s" protocol is not customizable).
3) Press
4) At the Clock menu (Display 17, page 41) the teal arrow represents a custom
protocol menu is available. H H H H old down
the 'READY' light (green LED) turns off.
5) At the New BT Device menu (Display 8 page 42) the teal arrow represents a
custom protocol menu is available. H H H H old down
seconds until the 'READY' light (green LED) turns off.
You will now see the Level L1 screen (Display 32, page 49). You are in the DPOAE menu and
will be able to scroll the available protocol parameters with the
changes by using the     LEFT or RIGHT
NOTE:
NOTE: If you push the
NOTE:
NOTE:
through date and time, etc., rather than accessing the custom protocol menus.
8106286-3_EROSCAN_US_15i.docx
Operation Manual ERO•SCAN
NOTE:
NOTE: When a change is made within the Advanced Options
NOTE:
NOTE:
menu to a default protocol, an * is place in the protocol name,
see Display 31.
CHANGE at the main menu (Display 2, page 23).
buttons, select the DPOAE protocol you want to
SETUP at the Protocol menu (Display 3, page 23).
arrow keys to CHANGE the selection.
DOWN arrow key without holding it for 3 seconds, you will scroll
48
old down
old down
old down
CHANGE key for 3 seconds until
old down the
old down
old down
NEXT button and make
®
CHANGE key for 3
11/15