3.1.2. Daily Test Procedure

 Phantom's value checking
Switch on the device and the main screen appears after booting up.
Select "setup" button to check the value of phantom.
When you see the bottom of the phantom, you can find a label which is two
kind of different label. You should check contents on the label and below two clauses.
1. the value of SOS and the range of BUA are correctly write on Phantom info.
2. If you can see 'V0.2' on the right side of label, you have to tick on the '0.2Ver'
radio box, if no, you must check the radio box. The box should be blank.
(If it is wrong with the information which comes from the label, you have to revise it.)
Check the Bottom
part of label
type 1
OT06-2R0423-06
DOC. No. : OT06-2R0423-06-MUL
type 2
21