User's Manual (Confidential)
OsteoSys Co., Ltd.
Model Designation: SONOST 3000
Chapter 3. Using SONOST 3000
3.1. Daily Test
The SONOST 3000 system is checkedusing the daily test.
The daily test should be performed at least once a day, before patient's measurements.
3.1.1. Preparation
SONOST 3000 device, Phantomand ultrasound transmission gel
Dailytest Recommendation
The daily test should be performedwithin operating temperature.
(Recommended temperature : 18 ~ 27 ℃)
The phantom should be kept beside of device closely.
Daily test should be done just after turning on the device.
Turn on the device and then do Dailytest immediately at beginning time
of day before warming up the device.
Turn off the device after using at closing time of day.
OT06-2R0423-06
DOC. No. : OT06-2R0423-06-MUL
20
A4(210 × 297)