Chapter 3. Using SONOST 3000

3.1. Daily Test

The SONOST 3000 system is checkedusing the daily test.
The daily test should be performed at least once a day, before patient's measurements.

3.1.1. Preparation

SONOST 3000 device, Phantomand ultrasound transmission gel
Dailytest Recommendation
 The daily test should be performedwithin operating temperature.
(Recommended temperature : 18 ~ 27 ℃)
 The phantom should be kept beside of device closely.
 Daily test should be done just after turning on the device.
 Turn on the device and then do Dailytest immediately at beginning time
of day before warming up the device.
 Turn off the device after using at closing time of day.