Using SONOST 3000

3.1 Daily test

The system of SONOST 3000 can be maintained by running Daily Test on a daily
basis.
The daily test should be performed at least once a day, before patient's
measurements.
3.1.1
Preparation
SONOST 3000 device, Phantom, and ultrasound transmission gel
NOTE! Daily Test Recommendation
The daily test should be performed within operating temperature.
(Recommended temperature: 18 ~ 27 ℃)
The phantom should be kept beside of device closely.
Daily test should be done just after turning on the device.
After turning on the device for the first time of the day, immediately
perform Daily Test immediately before warming up the device.
Turn off the device after using at closing time of the day.
3.1.2
daily test procedure
Phantom's value checking
Switch on the device and the main screen appears after booting up.
Select "setup" button to check the value of phantom.
When you see the bottom of the phantom, you can find a label which has
two different kinds of label. You should check the contents on the label and
below two clauses.
DOC. No. : OT06-2R7124