Characteristics - Keysight N2791A User Manual

25 mhz and 70 mhz high-voltage differential probes
Table of Contents

Advertisement

Characteristics

Characteristics for the N2791A and N2891A high-voltage differential probes are
shown below. The probe / oscilloscope should be warmed up for at least 20
minutes before any testing and the environmental conditions should not exceed
the probe's specified limits.
Electrical Characteristics
Bandwidth (-3 dB)
Attenuation ratio
Probe Risetime (10%-90%) 14 ns
Absolute Maximum Rated
Input Voltage (each side to
ground)
Maximum Differential Input
Voltage (DC + AC Peak)
Maximum Common Mode
Input Voltage
Input Impedance
Output Voltage Swing
Offset (typical)
CMRR (typical)
Power Requirements
Battery Life
Battery/Power Cord
- all are typical
N2791A
>25 MHz (driving 1 MW oscilloscope
input)
10:1, 100:1 (switchable)
1000 Vrms CAT II
\70 V at 10:1 attenuation
\700 V at 100:1 attenuation
\70 V at 10:1 attenuation
\700 V at 100:1 attenuation
4 MW, 10 pF (each side to ground)
8 MW, 8 pF (between inputs)
\7 V (driving 1 MW oscilloscope input) \7 V (driving 1 MW oscilloscope
\7.5 mV (adjustable)
-80 dB at 60 Hz, -40 dB at 1 MHz
4 AA batteries or USB power adapter
(5 V/200 mA, 9 V/120 mA)
15 hours (alkaline battery)
The supplied voltage must be less than
12 V and greater than 4.4 V or else the
probe could be damaged
N2891A
70 MHz (driving 1 MW oscilloscope
input)
100:1, 1000:1 (switchable)
5 ns
5000 Vrms CAT I
\700 V at 100:1 attenuation
\7000 V at 1000:1 attenuation
\7000 V at 10:1 attenuation
\7000 V at 1000:1 attenuation
50 MW, 7 pF (each side to ground)
100 MW, 5 pF (between inputs)
input)
\5 mV (adjustable)
-80 dB at 50 Hz, -60 dB at 20 kHz
4 AA batteries or USB power adapter
(5 V/200 mA, 9 V/120 mA)
11.5 hours (alkaline battery)
The supplied voltage must be less
than 12 V and greater than 4.4 V or
else the probe could be damaged
Characteristics
7