Offset Current Verification - Keithley 7072 Instruction Manual

Semiconductor matrix card
Table of Contents

Advertisement

SERVICE INFORMATION
Trim Cables
-A
707 Switching
Matrix
Figure 4-2. Connecting
the Test Cable to the Model 7072
Running
the Test
l
Model 617 Electrometer
l
Model 7078-TRX-3 Trim Cable
Follow the instructions
given in the Model 707 Instruction
Manual to perform the relay test. The computer will ad-
* Model 6172 2-&t
male to 3-lug female triaxial adapter
vise you as to which relay or group of relays (if any) fail
to p&s the test
Test Connections
4.3.4 Offset Current
Verification
Figure 4-3 shows the test connections
for offset current
verification.
The Model 7072 row being tested is to be con-
nected to the Model 617 Electrome&
input through
the
triaxial cable and the triaxial adapter. Note that the elec-
trometer
ground
strap is to be removed,
and the elec-
trometer should be operated in the unguarded
mode.
Recommended
Equipment
l
Model 707 Switching
Matrix
4-3