Fluke 8506A Instruction Manual page 281

Thermal true rms multimeter
Hide thumbs Also See for 8506A:
Table of Contents

Advertisement

The
8505
A
and
8506
A
use a nonvolatile, read /write
memory
for
microprocessor-
controlled
calibration of
each range
in
each
function.
The
Calibration
Memory
is
a
standard
part of
the
Controller
module
with
these multimeters.
Extended
intervals
between hardware
calibration are possible.
The
scale
factor for
each range can be
calibrated
using
any
reference
input value
from
60%
of
range
to
full
scale.
The
following
paragraphs
first
discuss the
general
procedure used
in
software
calibration,
then
detail
the
procedure used
for
each
function.
CAUTION
Interruption
of
input
power
could
affect
Calibration
Memory
entries
when
the
multimeter
Is
in
Calibration
mode.
Do
not cycle
Input
power
to
the
multimeter
when
Calibration
mod©
Is
activated.
If
power
Is
ON,
locally verify
that
the
AVG/(CAL)
annunciator
is
not flashing or remotely
verify
that
Calibration
mode
is
off
(G5 response
=
0)
before cycling
power
to
OFF.
If
power
Is
OFF,
verify
that
the
rear
panel
Calibration
switch
is
off
before cycling
power
to
ON.
Software
calibration
entails
a
combination
of zero
offset
and
gain
corrections.
The
Calibration
mode
must
be
activated
when making
calibration entries
from
either a local
or
remote
location.
This
mode
can
only be entered
by
accessing the Calibration switch
on
the
multimeter
rear panel.
The
calibration sticker
must
be
removed and
the
slide
switch
positioned
to
ON
when
the
multimeter
is
in
local (front
panel) control.
The
front
panel
AVG/(CAL)
annunciator
flashes to
denote
activation
of Calibration
mode.
Zero
correction values can then be entered
for
each range
in
dc
voltage
(VDC)
and
resistance
(OHMS)
functions
only.
Gain
correction factors
can be
stored
for
each range
in
each
function
by
applying a
reference input
and
entering
a
numeric
string
representing that
reference value.
The
multimeter then
computes
the gain correction factor necessary
to
read
the reference
value.
Depending on
the function being
calibrated,
gain corrections
are
made
once
or twice
for
each
range.
A
number
representing the calibration date or
identifying the
multimeter can
also
be entered
when
the
Calibration
mode
is
activated.
The
zero
offset
correction values are applied
to the
reading
whenever
the
Zero
mode
is
on
(ZERO
annunciator
lit).
The
gain correction
factors are applied to the
reading
as
soon
as
they
are entered
and
continue
to
be applied
whether
Calibration
mode
is
on
or
off.
NOTE
In Calibration
mode,
the
multimeter
uses
the
"permanent"
zero
offset
corrections stored
in
Calibration
Memory;
it
does not use
the
"temporary
zeros"
that
can be stored
when
Calibration
mode
is
off
Locally, application of zero correction values
can
be interrupted
by turning
the
Zero
mode
off.
Remotely, both
zero correction values
and
gain correction factors
can
be
inhibited
or enabled with
the
following
commands:

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents