Fluke 8506A Instruction Manual page 245

Thermal true rms multimeter
Hide thumbs Also See for 8506A:
Table of Contents

Advertisement

NOTE
The
instrument
must be connected
through a bus
network
to
a system
controller
,
e.g.,
the
Tektronix
4051
or
HP
9825,
to operate.
This troubleshooting
procedure
is
based
on
the
assumption
that the
instrument
has
been
checked
in
local
and found
to
be
operational
in
all
aspects prior to
installation
of the Bit
Serial
interface.
Install
the Bit
Serial
Interface
in
the instrument
and
apply
power from
the front panel switch.
Go
to the step
number
given
for
correct
Is
the
display
blank?
4
7
Check
the address
lines
and
address decoders.
Check
for
a
high
ACK
line.
Repair
as
required
and resume
at
step
2.
If
the display
is
incorrect (garbled or
wrong),
check
the input
ID
lines
and
gates.
Repair
as
required
and resume
at
step
2.
Using the
controller, instruct
the instrument to
go
to
remote (program
the character
"J").
Does
the
instrument go
into
remote?
Check
the input
gates
(TP1), the
UART
(U9), the
baud
rate at
TPS,
the
IWT
circuit,
and
the
status
output
buffer.
Program
several
instructions
from
the
remote
controller.
10
I
9
Does
the
instrument respond
correctly to
the
programmed
instructions?
13
12
Check
the
UART
(U9), the
output
gates
(U4),
and
the data
input
gate (U7). Repair
as
required
and resume
at step 10.
Troubleshooting
of the Bit
Serial Interface, as
applicable
at this
level,
is
complete.
606-5