Fluke 8506A Instruction Manual page 14

Thermal true rms multimeter
Hide thumbs Also See for 8506A:
Table of Contents

Advertisement

02A
Ohms
Converter
03
Current
Shunts
05
IEEE Standard
488-1
975
Interface
06
Bit
Serial
Asynchronous
'
07
Parallel
Interface
1}
Either
Option
02A
or
Option
03
can
be
installed
at
one
time.
2}
Only one
of
Options
05.
06.
and
07
a
M04-
205-600
S'A-inch
Rack
Adapter
MOO-260-61
0
18-inch
Rack
Slides
MOO-280-610
24-inch
Rack
Slides
80K-6
High Voltage
Probe
80K-40
83RF
85RF
Y8021
Y8022
Y8023
629170
High Frequency Probe
II
Y8022
IEEE
Std.
Cable,
2
Meter Length
Y8023
IEEE
Std.
Cable, 4
Meter Length
629170
TRMS
Extender
Card
MIS-71
90K*
Static
Controller
MIS-7013K*
Bus
interconnect
and
Monitor
*For use during
service or
repair.
ecmcations
GENERAL
SPECIFICATIONS
Dimensions
10.8
cm
High
x 43.2
cm
Wide
x 42.5
cm
Long
(4.25
in
High
x 17
in
Wide
x 16.75
in
Long) (See
Figure
1-1)
Weight
BASIC
....
10 kg
(22
lbs)
FULLY
LOADED
....
12 kg
(26
lbs)
Operating
Power
VOLTAGE
.
.
. .
100V
ac,
120V
ac,
220V
ac,
or
240V
ac
(±10%)
BASIC
INSTRUMENT POWER
.
.
.
.
.
.
12
watts
FULLY
LOADED POWER
24
watts
FREQUENCY
47
Hz
to
63
Hz
(400
Hz
available
on
request)
Warm-Up
.
.
.
.
2
hours
to
rated
accuracy
Shock and
Vibration
Meets requirements
of
MIL-T-28800
for
type
III,
class
5,
style
E equipment.
Temperature
Rang©
OPERATING
....
0°C
to
50°
NON-OPERATING
....
-40°
C
to
70°
Humidity
Range
C
TO
18°
C
....
80% RH
18°C
TO
40°
C
. .
.
.
75% RH
40°
CTO
50°
C
. .
. .
45% RH
Maximum
Terminal Voltage
LO
TO GUARD
....
127V rms
GUARD TO
CHASSIS
....
500V rms
H!
SENSE TO
Hi
SOURCE
127V rms
LO
SENSE TO LO SOURCE
....
127V rms
Hi
SENSE TO LO SENSE
1000V rms
or
1200V
dc
HI
SOURCE TO
LO
SOURCE
. .
.
280V rms
-2

Hide quick links: