Fluke 8506A Instruction Manual page 233

Thermal true rms multimeter
Hide thumbs Also See for 8506A:
Table of Contents

Advertisement

85G5A/8506A-05
Table
605-4.
Interface
Messages
1
ALL
DEVICES
RESPOND
(Universal)
ADDRESSED
DEVICES
ONLY
RESPOND
DEVICE
IN
LOCAL
RESPONDS
AND
GOES
TO
REMOTE
NOTE
BINARY
OCTAL HEX
i
My
Listen
Address
X
F
T AS A
4
A3 A2
A1
X
X
1
My
Talk
Address
X T
F
AS A4 A3 A2
A1
X
X
1
Uniisten
077
3F
X
UNT
Untalk
XTFTTTTT
137
5F
X
X
OTA
Other
Talk
Address
X
X
X X X
X X
X
X
2
SPE
Serial Poll
Enable
XFFTTFFF
030
18
X
X
SPD
Serial Poll
Disable
XFFTTFFT
031
19
X
X
LtO
Local
Lockout
XFFTFFFT
021
11
X
X
GTL
Go
To
Local
XFFFFFFT
001
01
X
DCL
Device Clear
XFFTFTFF
024
14
X
SDO
Selected Clear
Xppppypp
004
04
X
-
.
Illlf
.
605-25.
Resets
605-26.
Power-on
or
software
resets
may
occur.
At
power-on, U8(PB)-10
causes
an
interface reset to
prevent
unwanted
states
in
the interface
logic.
Software
resets,
decoded by
U13(PB)-6,
may
occur
as
a
result
of a
momentary power
interruption,
a
front
panel
request,
or
a system
request.
@05-27=
Control Register
605-28.
The
following
six
"D"
flip-flops
compose
the
control
register:
1.
Interrupt
enable
U14(PB)-2 remains
true,
except during
the
power-on
routine.
2.
A
service
request
(SRQ)
to the
system
controller
is
initiated
by U14(PB)-13.
3.
The
instrument
controller
being ready
for
data
(RFD)
is
indicated
by U16(PB)~I.
4.
At
the
last
data byte of
a
message
to
the
system
from
the
instrument
controller,
U11(PB)-13
goes
true.
5.
Data
accepted
(DAC)
is
sent
and
RFD
is
reset
by U16(PB-13.
6.
When
the
instrument
is
a
talker,
Ull(PB)-2
is
used
to
generate the data available
(DAV)
signal.
605-29.
Status Register
605-30.
The
status
register consists
of
U28
and
part
of
U31.
The
instrument address
decoded
by
U!3(PB)-9
enables a
status
byte
to
be
placed
on
the data
bus (IDO-
ID7).
This
status
byte
is
defined as
follows:
1.
IDO:
true
from U28-7
when
in
the
talk
only
mode.
2.
ID1:
true
from U28-9
when
remote
enable
(REN)
from
the
system
controller
is
false,
3.
ID2:
true
from U28-9
when
go
to local
(GTL)
is
true
from
the
message decoder
U5-4.
4.
IDS:
true
from U28-3
for
an
interface
message.
5.
ID4:
true
from U3I-3
for
an
interface
message.
6.
IDS:
true
from
U31-5
to
indicate a
device
dependent
message.
7.
ID6:
true
from U31-7
during
the
serial
poll
mode when
the
system
controller
is
requesting
status.
8.
ID7:
true
from U31-9
when
the
system
controller
is
requesting data
from
the instrument.
605-5