Hssl Failures/Message Reception Time-Out - CELL-DYN 3200 Service Manual

Automated hematology analyzer
Table of Contents

Advertisement

5#7URXEOHVKRRWLQJ •
IP-24 HSSL Failures/Message Reception Time-Out (Page 1 of 1)
CELL-DYN® 3200 SYSTEM SERVICE MANUAL
IP-24 HSSL Failures/Message Reception Time-Out (Page 1 of 1)
Start
- With Power OFF, check
and re-seat connectors on
the CPU/DCM, and Main
Computer.
- Re-seat the HSSL Board.
- Check and re-connect the
HSSL Cable.
- Reload HSSL Com
Problem
Corrected
?
Y E S
Total Call
- In order replace the
following.
- HSSL Cable
- CPU/DCM
- Serial Link/Keyboard
Controller
N O
(VP-67)
Part Number 9140300
(PL-17.3)
(PL-13)
(PL-17.2)
August 1998
2 - 77

Hide quick links: