VP-56 Background Counts Verification
Required Tools & Materials:
•
None
Verification Procedure:
1. From Main Menu, press
„ RUN
„ SPECIMEN TYPE
„ BACKGROUND
2. Press the "TOUCH PLATE" to run a background count.
3. Verify that the background counts are within specifications
according to the following table.
®
CELL-DYN
3200 SYSTEM SERVICE MANUAL
VP-56 Background Counts Verification
Table 5-7
Background Count Specification
WBC
NOC
(WOC)
µ
≤
≤ 0.10 k/
L
0.10 k/
VP-57 System CV Verification
Required Tools & Materials:
•
2.5 mL Fresh Normal Blood
Verification Procedure:
1. From Main Menu, press
„ RUN
„ SPECIMEN TYPE
2. Move the cursor to an empty file, and press
„ QC SPECIMEN
3. Mix the fresh blood thoroughly, and run 10 times in the
open mode and 10 times in the closed mode.
4. Press
„ MAIN
„ QUALITY CONTROL
„ VIEW QC LOG
5. Verify the CVs are within specification according to the
following table.
−
Part Number 9140300
August 1998
RBC
HGB
µ
µ
L ≤ 0.02 m/
L ≤ 0.1g/dL ≤ 5.0 k/
PLT
µ
L
5 - 53