Specifications; Electrical Specifications - AEMC SL261 User Manual

Ac/dc current oscilloscope probe
Hide thumbs Also See for SL261:
Table of Contents

Advertisement

3.1

Electrical Specifications

*Reference conditions: 23°C ± 5°C, 20 to 75% RH, DC to 1kHz, probe zeroed, 1 minute
warmup, batteries at 9V ± 0.1V, external magnetic field < 40 A/m, no DC component, no
external current carrying conductor, 1MΩ/100pF load, conductor centered in jaw.
Current Range:
100mV/A: 100mA to 10A peak
10mV/A: 1 to 100A peak
Output Signal: 1000mV peak max
AC Current Accuracy:
After calibration and for one year
(zero probe before making measurement)
Range
100mV/A
(50mA to 10A
10mV/A
(500mA to 40A
10mV/A
(40A to 80A
DC/AC
10mV/A
(80A to 100A
Frequency Range: DC to 100kHz (-3dB with current derating)
Noise:
Range 10mV/A: 480µV
Range 100mV/A: 3mV
Slew Rate:
Range 10mV/A: 20mV/µs
Range 100mV/A: 0.3V/µs
Load Impedance: > 1MΩ/100pF
AC/DC Current Oscilloscope Probe Model SL261

SPECIFICATIONS

± 3% of reading ± 50mA < 1.5° from DC to 65Hz
peak)
DC/AC
± 4% of reading ± 50mA
peak)
DC/AC
peak)
peak)
DC/AC
CHAPTER 3
Accuracy
12% of reading ± 50mA
15% of reading
Phase Shift
< 1° from DC to 65Hz