Fig. 22 Mounting Of Intermediate Tube; Fig. 23 Back Of Stand - Zeiss Axioplan 2 Operating Manual

Hide thumbs Also See for Axioplan 2:
Table of Contents

Advertisement

Microscope Components
5
4
3
2
1
Fig. 22
Mounting of intermediate tube
Fig. 23
Back of stand
11 12
B 40 042 e 09/99
General
Intermediate tubes are always attached to the upper part of
the stand. To do this, the binocular tube must first be removed
and its tube lens removed.
6
7
Note:
tube lens (4).
8
Mounting of an intermediate tube

9









1
Use the cover of the tube lens container to remove the
Switch off your instrument.
Loosen clamping screw (2) on stand (1) and remove the
tube (5) upwards.
Attach the intermediate tube (8) to the stand.
Centration is made using the dovetail.
Tighten clamping screw (2) on stand only slightly at first.
Insert the supplied clamping screw with washer (7) into
the drilled hole provided and screw it into the threaded
hole (9) of the stand.
Align the intermediate tube with the edges of the inter
mediate tube and screw tight clamping screws (2) and (7).
Remove the tube lens (4) from the tube because a tube
lens is integrated in the intermediate tube.
Attach the tube (5) to the intermediate tube (8) and
center via the dovetail.
Align the tube with the edges of the intermediate tube
and tighten the clamping screw (3).
For coded or motorized intermediate tubes, connect the
serial interface of the intermediate tube (Fig. 22/6) with
that of the stand (Fig. 23/1) via the RS 232 C serial cable
supplied.