Unit Settings [MS]
M
/ M
AX
IN
With this function you can select if the maximum
or minimum value will be shown permanently in
the upper part of the display.
To switch between both please press the M
button, either during the HOLD mode or during
a measurement (while the T
The made setting will be saved, also after the
unit switched off.
If you press the M
measurement (during the HOLD mode) the
determined maximum and minimum value taken
during that measurement will be displayed.
E
MISSIVITY
The intensity of infrared radiation, which is
emitted by each body, depends on the
temperature as well as on the radiation features
of the surface of the measuring object.
The emissivity ( = Epsilon) is used as a stable
factor of the material, with which to describe the
ability of the body to emit infrared energy. If the
emissivity chosen is too high, the infrared
thermometer may display a temperature value
which is much lower than the real temperature.
The unit will be delivered with a preset fixed
emissivity of 0,95. This emissivity value is very
common for most organic materials and painted
or oxidized surfaces.
Shiny or metallic surfaces may result in
inaccurate reading due to reflexions. To prevent
this, cover the measuring surface with either flat
black paint or with plastic labels.
°C/ °F S
ETTING
To setup the temperature unit to °F please
press the D
the T
RIGGER
To setup the temperature unit to °C please
press the U
the T
RIGGER
ODE
button (keep pressed) and then
OWN
.
button (keep pressed) and then
P
.
optris MS – DE2011-04-A
RIGGER
button after you made a
28
ODE
is pressed).