1
Introduction to the Variable Wavelength Detector
Special Settings
The special settings are required for specific actions (normally in a service
case).
The tables include both settings for modules – with on-board LAN and without on-board
N O T E
LAN. They are identified as LAN and no LAN.
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident mode. It is not operable as a
module. It only uses basic functions of the operating system for example, for
communication. In this mode the main firmware can be loaded (using update
utilities).
Table 11
Boot Resident Settings (without on-board LAN)
Mode Select
LAN
TEST/BOOT
No LAN
TEST/BOOT
34
SW1
SW2
SW3
1
1
1
1
1
0
Agilent 1260 Infinity Variable Wavelength Detector User Manual
SW4
SW5
SW6
0
0
0
0
1
0
SW7
SW8
0
0
0
0