8
Test Functions
Filter/Grating Motor Test with Instant Pilot
The test can be started via the command line.
• To start the test
• Test succeeded (when passed)
Grating Motor Test Failed
Probable cause
1 Defective grating motor
2 Defective grating position sensor
3 Defective VWM board
Filter Motor Test Failed
Probable cause
1 Defective filter motor
2 Defective filter position sensor
3 Defective VWM board
132
TEST:GRAT or
TEST:FILT
Reply: RA 00000 TEST:GRAT or RA 00000 TEST:FILT
Agilent 1260 Infinity Variable Wavelength Detector User Manual
Suggested actions
Exchange Optical Unit.
Exchange Optical Unit.
Exchange the VWM board
Suggested actions
Exchange Optical Unit.
Exchange Optical Unit.
Exchange the VWM board