Exchanging the Outlet Valve
When
if leaking internally
Tools required
Description
Wrench, 1/4 - 1/5 inch
Wrench 1/4 inch
Wrench, 14 mm
Parts required
p/n
G1312-60067
Preparations
Switch off the pump at the main power switch
1 Using a ¼ inch wrench disconnect the absorber capillary from the
2 Unscrew the valve with the 14 mm wrench and remove it from the
3 Do not disassemble the outlet valve, as this can damage the valve.
4 Reinstall the outlet valve and tighten it using a torque wrench (12 Nm).
1260 Infinity Binary Pump User Manual
Description
Outlet valve 1220/1260
outlet valve.
pump body.
Maintenance
Exchanging the Outlet Valve
10