Setting up the Agilent 1260 Infinity Binary Pump G1312B with the Instrument Control Interface

Pump Control

The pump can be switched between following states: On, Off or to Standby.
In Standby, the pump motor is still controlled. When the pump is switched
on from standby, it does not re- initialize.
Upon initialization, the pump ignores the Maximum Flow Gradient value.
C A U T I O N
This can result in a rapid and uncontrolled pressure increase.
The optional seal wash pump can be controlled by either switching it off,
using it for a single time or specifying frequency and duration of periodic
wash intervals.
1260 Infinity Binary Pump User Manual
To prevent harm to the column, open the purge valve until the initialization is
finished.