Setting up the Agilent 1260 Infinity Binary Pump G1312B with the Instrument Control Interface

Data Curves

The binary pump provides the possibility to store the following operational
data in the data file of the Agilent data system:
• Solvent percentage for each channel,
• pump flow,
• pressure
The pressure data curve is generated from the pressure sensor readings, while %A, %B and
N O T E
flow are calculated from the method settings of the pump.
For details, please refer to the online help or manual of your instrument
control software.

Bottle Filling

The pump offers a powerful feature to monitor the liquid level in the
solvent bottles. With total bottle volume and initial filling volume set
correctly, the pump subtracts the consumed volume continuously from the
initial value and stops the pump and method/sequence execution before
the system runs dry or an analysis is corrupted.
The bottle filling feature fails if multiple solvent inlets are put into one solvent bottle!
C A U T I O N
Table 6
1260 Infinity Binary Pump User Manual
In that case implement a minimum pressure limit (see
that the pump runs dry when solvents are empty.
on page 68 lists the available bottle filling parameters.
Using the Pump
Table 4
on page 64) to avoid
4