Servo Restart Failed

Error ID: 2201, 2211
Servo Restart Failed: Pump channel A
B: Servo Restart Failed: Pump channel B
The pump motor in the module was unable to move into the correct
position for restarting.
When the module is switched on, the first step is to switch on the C
phase of the variable reluctance motor. The rotor should move to one of
the C positions. The C position is required for the servo to be able to take
control of the phase sequencing with the commutator. If the rotor is
unable to move, or if the C position cannot be reached, the error message
is generated.
Probable cause
1
2
3
4
5
118
Disconnected or defective cable.
Blocked active inlet valve.
Mechanical blockage of the module.
Defective pump drive assembly.
Defective main board.
Suggested actions
Please contact your Agilent service
representative.
Exchange the active inlet valve.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.