iv
VisualSonics Vevo Compact Anesthesia System Operator Manual
Need help? Call us toll-free: 1-866-416-4636 or E-mail us: support@visualsonics.com
Keep the vaporizer upright at all times.
Never tilt a calibrated vaporizer beyond 45 degrees when charged.
Dangerous overdose may result during subsequent use. This
condition can be detected by calibration verification only.
Do not connect the vaporizer directly into the patient breathing
circuit.
Do not turn on two vaporizers at the same time.
Do not put water or any other solvent in a vaporizer.
Due to the wick's capacity to absorb agent, the vaporizer will
continue to deliver the set concentration for a considerable period of
time even if no liquid level can be viewed through the sight window.
However, do not use a vaporizer in this condition on a subject.
Flush flows through an open vaporizer may result in unwanted
concentrations of anesthetic agent.
Do not ignore maintenance of your vaporizer. An annual calibration
recertification is highly recommended by the manufacturer.
CAUTIONS
Personnel operating this compact anesthesia system must be
thoroughly familiar with the instruction manual and equipment
operation prior to use with subjects.
Tampering with the anesthesia system components by unauthorized
personnel voids all warranties and specifications. The prevention of
tampering with the anesthesia system is the sole responsibility of the
user or owner. VisualSonics assumes no liability for any malfunction,
failure, damage or loss to either equipment or life.
Before using the anesthesia system check that it is set up properly,
ensure all connections and settings are correct.
Always adhere to proper Diameter Index Safety System (D.I.S.S.)
hose connections. Failure to comply will void your warranty and
could compromise subject safety.
Remove the machine from service if:
a.
the oxygen flush button does not immediately return to the OFF
position after being released, or
b. the flow of oxygen does not immediately stop flowing from the
common outlet after the oxygen flush button is released.
Then call VisualSonics or a VisualSonics-authorized service
facility to repair the valve.
If the oxygen flush button does not immediately return to the off
position after being released, or if the flow of oxygen does not
immediately stop flowing from the common outlet after the oxygen
V1.0