Chapter 2: Installing the System: Troubleshooting
Troubleshooting
You hear a hissing sound
14
VisualSonics Vevo Compact Anesthesia System Operator Manual
Need help? Call us toll-free: 1-866-416-4636 or E-mail us: support@visualsonics.com
Connect the oxygen supply to the mobile vaporizer sub-system or the
2.
table-top vaporizer sub-system as described in "To connect the oxygen
supply to the vaporizer sub-system:" on page 11.
This indicates that there is a leak in the cylinder valve, regulator to
cylinder connection, or in the regulator.
To stop the leak:
Close the cylinder valve and open the regulator momentarily to
1.
relieve all pressure.
Return the regulator to the off position.
2.
Re-tighten the regulator to the cylinder connection and repeat the
3.
test.
If the hissing sound continues it means that one or more of the following
conditions exist:
•
The sealing washer is leaking and should be replaced. Make sure the
cylinder valve is fully turned off before removing the regulator.
•
The cylinder valve is leaking. Place the cylinder outdoors and notify
the gas supplier immediately. Do not attempt to use the cylinder.
•
The regulator is leaking and must be removed from service and
returned for repair. Remember to always turn the cylinder valve fully
off before removing or loosening the regulator connection.
V1.0
Need help?
Do you have a question about the VEVO and is the answer not in the manual?