Chapter 6: Service
Maintenance
Vaporizer
Stand Assembly
Flowmeter(s)
Oxygen Flush
Rubber Goods
Equipment
V1.0
VisualSonics Vevo Compact Anesthesia System Operator Manual
Need help? Call us toll-free: 1-866-416-4636 or E-mail us: support@visualsonics.com
Annual preventive maintenance and certification by qualified personnel
is recommended to ensure proper operation. General upkeep and daily
maintenance will enhance the longevity of your compact anesthesia
system.
Warning: Do not modify, tamper with or disassemble
the vaporizer. Doing so could damage the unit and
alter the accuracy of graduation.
Observation of the operating instructions, regular servicing and normal
professional vigilance is normally all that is required to maintain the
vaporizer in a safe working condition.
Cleaning
Clean the exterior of the vaporizer with a damp cloth.
Never allow cleaning agents to accumulate in the filler, gas inlet and
outlet ports, or around the control dial.
Do not use harsh or abrasive cleaners. To clean, wipe with a damp cloth.
The stand is coated with a baked-on enamel that will withstand most
cleaning/disinfectant solutions. If a solution is to be used, it is
recommended that you initially test its effect on an inconspicuous spot of
the enamel.
Do not use harsh or abrasive cleaners. Plastic flowmeter covers may be
removed and cleaned with a damp cloth. The exterior of the glass
flowtube may be cleaned in the same manner. Should dirt or an
obstruction get inside the flowtube, contact VisualSonics for assistance.
The flow control requires no maintenance.
This valve requires no maintenance.
Rubber is a natural product and will deteriorate with age and use. All
rubber goods are considered disposable and should be inspected daily for
presence, condition and alignment. DO NOT USE PETROLEUM BASED
LUBRICANTS. If you must use a lubricant, use only a silicone-based
product.
This equipment must be periodically checked, calibrated, maintained
and, components repaired and replaced when necessary for equipment to
operate reliably. Parts that have failed in whole or in part, exhibit
excessive wear, are contaminated or are otherwise unfit for use, should be
immediately discarded and replaced. To maintain the warranty,
replacement parts must be installed or approved by VisualSonics Inc.
Equipment that is not functioning correctly must not be used. This
equipment must not be modified by unauthorized personnel or with
unauthorized components.
Chapter 6: Service: Maintenance
27